Radiation Oncology Database Server Requirements (5U Rack Mounted)

Processor:
Dual Intel® Quad Core Xeon Processor with 2x6MB Cache, 2.66GHz, 1333MHz FSB or better
Bus Type:
PCI-E
Memory (DRAM):
4 GB Fully Buffered DIMM's at 667MHz Dual Ranked
Hard Drives:
4x146 GB Hot Swap drives with SAS interface (includes online spare). Drives should have < 10ms access time at 15000 RPM. Effective disk space is 292 GB.
Drive Controller:
Dual channel SAS 3.0Gb/s RAID controller with ≥ 256 MB cache. Configured for hardware RAID 5 fault tolerance and Read/Write cache enabled (software NOS controlled RAID 5 not recommended). CD and LTO tape should operate from channel 2 of controller or a separate controller.
Floppy Disk:
3.5" drive, 1.44 MB
Optical Drive:
Multisession 24X CD-RW and DVD-ROM combo Drive
Backup Tape:
LTO-4-120 800/1.6TB internal tape drive; BrightStor™ ARCserve® 11.5 or later.
Network Interface:
1 LOM 10/100/1000Mbps, compatible with the network operating system and cabling.
Redundancy:
PSU and Fans
Ports:
1 serial, minimum 2 USB
Operating System:
ARIA: Microsoft® Windows® 2003 R2 Server (Standard or Enterprise) with Service Pack2 (25 client access license or sufficient seat licenses)
File System:
NTFS
Microsoft® OS Architecture
32-bit (not 64-bit)
Microsoft® .NET Framework
ARIA(8.0 ≥ 8.5): 1.1 and 2.0 (not 3.0)
ARIA(8.6): 1.1, 2.0 and 3.0
42U Half Height Rack systems come with the following components (see notes below):
  • Cables for Keyboard and Video
  • NEMA standard power strip
  • 1U 8-port keyboard/monitor KVM switch box
  • 1U keyboard/trackball and monitor console
  • 2U 3000VA Uninterruptible Power Supply (UPS)
  • 24-port patch panel (CAT5E)

Notes:

  1. Hardware used should be approved based on the current Microsoft® Windows® Catalog Listing.
  2. In addition to this Database Server, the following Image Fileserver is required for use with all image-based products and would be installed in the same rack.
  3. The Operating System (build and version) on the Database Server shall match the Operating System (build and version) on the HARRP server.
  4. 1 Gbps network recommended for maximum speed and performance.
  5. Specifications subject to change without notice as newer computer models become available.

 

Contact Varian Oncology

Varian Oncology Headquarters
Tel: 1.650.424.5700
Contact Us

Events

  • 11/18/2008 - 11/21/2008
    Medica
  • 11/26/2008 - 11/29/2008
    AMPI Conference
  • 11/30/2008 - 12/05/2008
    RSNA 2008
  • 12/06/2008 - 12/09/2008
    ASH 2008 Annual Meeting (American Society of Hematology)

© 1999-2008 Varian Medical Systems, Inc. All rights reserved.
Production of any of the material contained herein in any format or media without the express written permission of Varian Medical Systems is prohibited.

North America EMEA - Europe, Middle East, India & Africa